DE LAMA – Autoclave – DLOV PLSA
Sold
Specifications
Manufacturer | De Lama |
Model | DLOV PLSA |
Year | 1999 |
Condition | Used |
Serial Number | MLTC-IT346 |
Description
Supply voltage: 380 V, 50 Hz, three phase
Installed power: 4,5 kW
Maximum working pressure: 2.5 bar, 250 kPa
Maximum operating temperature: 138.2 °c, 411.2 K
Useful capacity: 1000 liters
Internal dimensions: height 1000 mm, width 800 mm, depth 1250 mm.
External dimensions: height 2150 mm, width 1700 mm, depth 1800 mm.
Theoretical empty weight: 2100 kg
Theoretical weight at hydraulic testing: 3100 kg
Theoretical working weight (excluding product): 2150 kg
Water supply vacuum pump:
connection size: Ø ½ “” GAS UNI338
Minimum pressure: 1.5 bar
Maximum temperature: 20 °C
Maximum required flow rate: 0.9 m3 / h (15 liters per minute)
Indicative consumption per cycle: 750 litres
Water supply for cooling:
connection size: Ø ½ “” GAS
Minimum pressure: 2.5 bar
Maximum temperature: 20 °C
Maximum required flow rate: 0.72 m3 / h (12 liters per minute)
Clean dry saturated steam:
connection size: Ø ¾ “” DN 20
maximum pressure: 3.5 bar, minimum pressure: 2.5 bar
maximum capacity required: 90 Kg / h
indicative consumption cycle: 68 Kg
industrial steam for interlocks:
connection size: Ø ½ “” GAS UNI338
maximum pressure: 3.5 bar, minimum pressure: 2.5 bar
maximum capacity required: 45 Kg / h
indicative consumption cycle: 34 Kg
sterile compressed air room entrance:
connection size: Ø ½ “” GAS UNI338
maximum pressure: 3.5 bar, minimum pressure: 2.5 bar
Maximum capacity required: 50 m3/h (840 liters per minute)
indicative consumption cycle: 16.800 liters
compressed air tools:
for controls – connection dimensions: Ø ½ “” GAS UNI338
maximum pressure: 7 bar, minimum pressure: 6 bar
Maximum capacity required: 5 m3 / h (83 litres per minute)
Electrical power installed:
system: three phase + ground + neutral
voltage: 380V-frequency: 50Hz
power: 3.55 kW
superpres discharge. Chamber safety valve:
connection size: 1 ½ “” GAS F UNI338
Maximum temperature: 140 °C
superpres discharge. Safety valve intercap.:
connection size: 1 ” GAS F UNI338
Maximum temperature: 140 °C
Vacuum pump discharge-total discharge:
connection size: 1 ½ “” GAS UNI338
Maximum temperature: 110 °C